Blood Pressure Device Accuracy Evaluation: Statistical Considerations

A comprehensive statistical analysis of the accuracy of blood pressure devices based on the method of AAMI/ANSI SP10 standards developed by the AAMI Sphygmomanometer Committee for indirect measurement of blood pressure, incorporated into IS0 81060-2. The 'bpAcc' package gives the exact probability 'of accepting a device D' derived from the join distribution of the sample standard deviation and a non-linear transformation of the sample mean for a specified sample size introduced by Chandel et al. (2023) and by the Association for the Advancement of Medical Instrumentation (2003, ISBN:1-57020-183-8).